Revision B MAC 5000 resting ECG analysis system 5-5
2024917-010
Troubleshooting: General Fault Isolation
Power-up Flow Chart
Poor Quality ECGs
Poor quality ECGs can be caused by factors in the environment,
inadequate patient preparation, hardware failures related to the
acquisition module, leadwires, cables, or problems in the unit.
Use a simulator to obtain an ECG report. If the report is good, the
problem is external to the unit.
84A