Revision B MAC 5000 resting ECG analysis system 4-19
2024917-010
Maintenance: Disassembly Guidelines
Restore Options
Using the option activation codes for the system, restore the options which had been
installed on the board which was removed. These options are printed on a label
located on the bottom of the paper tray as shown below.
Location of Option Activator Code Label
NOTE
Use the activator codes provided on the label found on your system.
The activator codes shown in the figure are provided as examples
only.
1. Power on the cart.
2. Within the System Setup function, select Basic System.
3. Select Option Activation to activate options.
4. Type the 12-digit option activation code and press the Enter key.
5. Repeat the previous step for each option to be activated on the new system.
6. Highlight Return and press Enter to return to the Basic System
menu.