Revision B MAC™ 5000 resting ECG analysis system 4-5
2024917-010
Maintenance: Inspection and Cleaning
Thermal Printhead
Clean the thermal printhead every three months or more often with
heavy use. A build-up of thermal paper coating on the printhead can
cause light or uneven printing.
Use a solution containing alcohol on a nonwoven, nonabrasive cloth such
as Techni-Cloth to wipe off the printhead. Do not use paper toweling, as
it can scratch the printhead.
Thermal
Printhead
35A