Revision B MAC™ 5000 resting ECG analysis system 4-3
2024917-010
Maintenance: Introduction
Introduction
Recommended Maintenance
Regular maintenance, irrespective of usage, is essential to ensure that
the equipment will always be functional when required.
WARNING
Failure on the part of all responsible individuals,
hospitals or institutions, employing the use of this device,
to implement the recommended maintenance schedule
may cause equipment failure and possible health
hazards. The manufacturer does not in any manner,
assume the responsibility for performing the
recommended maintenance schedule, unless an
Equipment Maintenance Agreement exists. The sole
responsibility rests with the individuals, hospitals, or
institutions utilizing the device.
Required Tools and Supplies
In addition to a standard set of hand tools, you will need the items listed
below.
Table 1. Tools and Supplies
Item Part Number
#10 TORX driver
Leakage current tester MT-1216-02AAMI (for 220V)
MT-1216-01AAMI (for 110V)
Multifunction micro-simulator MARQ 1
Precision dust remover
Lint-free soft cloth TX609
PS2 style keyboard (Japan only)