Revision B MAC 5000 resting ECG analysis system 3-17
2024917-010
Installation: MAC 5000 ST Requirements and Configuration
Configure pin 2 on the ANA/TTL port by selecting the following:
System Setup,
Exercise Test,
Inputs/Outputs, and
set Slow Analog Output to Workload.
Configure pin 8 on the ANA/TTL port by selecting the following:
System Setup,
Exercise Test,
Inputs/Outputs, and
Set Fast Analog Output to Workload.
Bicycle Ergometers
Ergoline 800/900, Lode Ergometer
Connection Requirements – Use cable PN 2008109-001 (Ergoline 800),
PN 2008114-001 (Ergoline 900), or PN 2007981-001 (Lode Ergometer), to
connect from the MAC 5000 ANA/TTL port to the ergometer analog
control port.
NOTE
For any other ergometer, the customer is responsible for making the
appropriate cable.
Device Configuration Requirements – Refer to ergometer Operator's
Manual.
MAC 5000 Configuration Requirements – Use the Edit Protocol
application to set the protocol Test Type to Ergometer in Watts or
Ergometer in KPM for protocols that will be used with this ergometer.
Configure pin 2 on the ANA/TTL port by selecting the following:
System Setup,
Exercise Test,
Inputs/Outputs, and
Slow Analog Output to Workload, or
Configure pin 8 by selecting:
System Setup Exercise Test Inputs/Outputs
Fast Analog Output to Workload

Table of Contents

Other manuals for GE MAC 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5000 and is the answer not in the manual?

GE MAC 5000 Specifications

General IconGeneral
BrandGE
ModelMAC 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals