Troubleshooting
Message Cause Solution
Battery not detected The battery is not detected by
the device.
Perform below steps:
1. Remove the battery.
2. Re-insert the battery.
If the battery is still not
detected, replace the battery.
Contact your GE Healthcare
Service Support representative
if the error persists.
Device date/time is incorrect.
Update.
The date and time set on the
device is not correct.
Select Adjust to set the correct
date and time. For more
information, see Configuring
Date and Time chapter of MAC
VU360
Resting ECG Analysis
System Setup and Configuration
Manual.
Keyboard failure The keyboard is not working. Contact your GE Healthcare
Service Support representative
if the error persists.
Touchscreen failure The touchscreen is not working. Contact your GE Healthcare
Service Support representative
if the error persists.
Cannot perform action while
acquiring ECG data
You have tried to perform
below actions while acquiring
an ECG or rhythm:
Start a test for a new
patient
Enter or edit patient
information (available for
rhythm)
Change speed, gain, or filter
(available for rhythm)
Access User Menu
Access Orders, Files, or
Queue list
Start ECG
Power off, lock, log out,
standby or privacy mode
Perform only allowed actions.
ECG Acquisition Errors
The table lists messages that you may encounter while acquiring ECG.
122 MAC VU360
Resting ECG Analysis System 2088531-370-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
BrandGE
ModelMAC VU360
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals