Patient Information
1. Start a test for a new patient. For more information, see Start a Test for a New
Patient on page 31.
2. Scan the patient’s barcode to populate the Patient Information screen.
Position the barcode reader 10 cm to 15 cm (4 inches to 6 inches) above the
barcode to be scanned.
If you are using the Acquisition Module barcode reader, press the Barcode Scan
button.
If you are using an external barcode reader, the barcode is automatically
scanned.
The barcode reader emits a sound to confirm the barcode scan. The Patient
Information screen expands and displays the fields populated with patient
information.
3. If there is a mismatch between the data scanned from the barcode and existing
patient information, a message displays. Perform one of the steps as follows:
• Select Use Scanned Data to populate the barcode data into the relevant
fields on the Patient Information screen, and confirm the data entered from
the barcode is accurate.
• Select Use Current Data to retain the manually entered information on the
Patient Information screen, and enter or modify patient information as
necessary.
Query Orders or ADT for Patient Demographics
Make sure that:
• You have privileges to see orders and do remote patient query.
• Use orders of ADT data as configured on the device to do remote patient query.
• A barcode reader is connected to the device.
1. Start a test for a new patient. For more information, see Start a Test for a New
Patient on page 31.
2088531-370-2 MAC VU360
™
Resting ECG Analysis System 33