Technical Specifications
ECG Specifications
Table 59: ECG Data Acquisition Specifications
Item Description
Signal input Type CF, Defibrillation-Proof
Defibrillation protection: Per IEC 60601-2-25:2011
Dynamic range AC Differential ± 5mV, DC offset ±300 mV
Common mode rejection >130 dB (>100 dB with AC filter disabled)
Input impedance >10MΩ @ 10 Hz, defibrillator protected
Patient leakage current <10 µA
Detection of pacemaker pulse Duration: 0.2 ms to 2.2 ms
Amplitude: 2 mV to 700 mV
Separation: 1 ms or greater
Pace Annotation Dedicated pace channel on display and printed reports
Pace digital sampling rate 75,000 samples/second per channel
Table 60: ECG Data Processing Specifications
Item Description
ECG Interpretation Marquette 12SL ECG Analysis Program for Adults and Pediatrics
Computerized measurements 15-lead analysis (includes measurements of user-selectable
additional 3 leads)
Additional ECG function Vectorcardiography
Heart rate meter 30 to 300 beats per minute (BPM) with an accuracy of ±10% or 5
BPM, whichever is greater. Heart rates outside this range will not
display.
ECG data formats GE Hi-Fidelity ECG, XML
Pre-acquisition Provides 10 seconds of instantaneous ECG acquisition
Digital rhythm Up to 5 minutes of continuous rhythm storage (exportable as a
PDF)
Storage 1000 records consisting of 10 second Resting ECG records and
Digital Rhythm records on the device internal memory
External storage USB-compliant flash drive supporting the FAT32 file system
2088531-370-2 MAC VU360
™
Resting ECG Analysis System 163