Troubleshooting
Message Cause Solution
Communication Error with
Acquisition Module
The acquisition module is
unable to communicate with
the device.
The acquisition module
recovers from the error unless it
is a hardware failure.
If the error occurs while the
device is recording or printing
an ECG or rhythm report, the
operation is aborted. Restart
the recording or printing
manually.
Report generation failed
Cannot open report
ECG recording failed
Unknown error 1. Retry the action.
2. If the error persists, power
off and power on the
device.
3. If the error persists,
contact your GE
Healthcare Service
Support representative.
Printing Errors
Table 29: Printer Errors Encountered During ECG Patient Report Printing
Error Message Error Condition How to troubleshoot:
Printer error. Paper jam
detected.
Paper is jammed. Carefully remove the stuck
paper from the rollers inside
the printer and verify that the
remaining papers are loaded
correctly in the tray. Printing will
restart automatically.
Printer error. Out of paper. The paper tray is empty. Insert sufficient paper in the
paper tray. Printing will restart
automatically.
Printer error. Door is open. The printer door is open. Close the printer door. Printing
will restart automatically.
2088531-370-2 MAC VU360
Resting ECG Analysis System 123