GE COMPANY
DIRECTION 5472001-1EN, REVISION 6OPTIMA CT680 SERIES AND OPTIMA CT670 INSTALLATION MANUAL
Page 138 Section 3.0 - Shim Installation
Section 3.0 Shim Installation
3.1 Time and Personnel
3.2 Tools and Test Equipment
Standard FE Tool Kit
•Shim Kit
NOTICE Understand and Follow All General Table Safety Procedures.
3.3 Preparation
Check head holder for a tight fit. If the head holder fit is loose, follow this procedure and shim for.
Axial head holder
Foot extender
Phantom holder
Introduction:
Some Axial Head Holders have a large free-play in the horizontal direction which could
potentially lead to motion and therefore image artifacts.
Installation of the 2327335 rubber shim kit can minimize this motion.
Figure 3-3 Axial Head Holder
Notes before Selecting Shim Thickness:
While selecting the best shim size, do not attach the rubber shim yet using the adhesive on the
back. It is best to use a piece of tape to hold on the shim in order to see if the size is correct.
Selecting a shim size that is too thick may result in:
- Difficulty latching the head holder properly. The head holder must latch so that a patient
is not injured.
Required Persons Preliminary
Reqs
Procedure Finalization
1 (FE or mechanical supplier) 10 min. 15 minutes labor on-site 5 min.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Optima CT680 Series and is the answer not in the manual?

GE Optima CT680 Series Specifications

General IconGeneral
BrandGE
ModelOptima CT680 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals