Page 62 Section 9.0 - Table Installation (GT1700V)
9.3.7 Tighten the Lock Rings
1.) Re-check gantry bubble levels.
2.) Re-check that each of the eight adjuster is loaded by attempting to turn it.
CAUTION Eye protection is required when using a hammer and chisel.
3.) Tighten the lock rings at all locations with the spanner, where possible. Use a hammer and
chisel to tighten the lock rings only where you can not use the spanner.