Probe
Two depth ranges full and half depth range
Very fast (approximately 0.1 s) response allows measuring from a vehicle
Up to 10 calibrations, up to 6 of them made by user
Measured quantities:
o Apparent conductivity in mS/m
o In-phase ratio in ppt (determined by magnetic susceptibility)
Measuring ranges:
o Conductivity: 1000 mS/m, resolution 0.1 mS/m
o In-phase ratio: ±80 ppt, resolution 10 ppm,
±400 ppt for CMD-Duo, informative
Measurement accuracy: ± 4% at 50 mS/m
Temperature stability: ±0.1 mS/m / °C
Effective depth range for High / Low depth mode:
o CMD-Tiny: 0.7 m / 0.35 m
o CMD-1: 1.5 m / 0.75 m
o CMD-2: 3.0 m / 1.5 m
o CMD-4: 6.0 m / 3.0 m
o CMD-4/6: 6.0 m / 3.0 m; 9.0 m / 4.5 m
o CMD-Explorer 2.2 4.2 6.7 m / 1.1 2.1 3.3 m
o CMD-MiniExplorer 0.5 1.0 1.8 m / 0.25 0.5 0.9 m
o CMD-Duo 15 30 60 m / 7,5 15 30 m
Inter-coil spacing:
o CMD-Tiny: 0.45 m
o CMD-1: 0.98 m
o CMD-2: 1.89 m