EasyManuals Logo
Home>Honeywell>Gas Detectors>MultiRAE Series

Honeywell MultiRAE Series User Manual

Honeywell MultiRAE Series
195 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
Alarm Signal Summary
Hygiene Mode
Alarm Type
Buzzer &
LED
Display Vibration Reading Backlight Priority
Super Alarm
4
beeps/sec
“Super
Alarm”
screen
400ms - On Highest
Man Down
Alarm
3
beeps/sec
“Man Down
Alarm”
screen
400ms - On
Man Down
Warning
2
beeps/sec
“Are you OK”
screen
400ms - On
Fail
3
beeps/sec
“Lamp” at
PID location
“Off” at LEL
location
400ms
Blinking
reading
On
Pump
3
beeps/sec
Blinking
pump
symbol
400ms Reading On
Max
3
beeps/sec
“Max” at
sensor
location
400ms
Blinking
reading
On
Over Range
3
beeps/sec
“Over” at
sensor
location
400ms
Blinking
9999
On
High
3
beeps/sec
“High” at
sensor
location
400ms Reading On
Low
2
beeps/sec
“Low” at
sensor
location
400ms Reading On
Negative
1
beep/sec
“Neg” at
sensor
location
400ms 0 On
STEL
1
beep/sec
“STEL” at
sensor
location
400ms Reading On
TWA
1
beep/sec
“TWA” at
sensor
location
400ms Reading On
Calibration
Fail
1
beep/sec
“Cal” at
sensor
location
400ms Reading On
Bump Fail
1
beep/sec
“Bump” at
sensor
location
400ms Reading On
Honeywell MultiRAE Series™ 150 User Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell MultiRAE Series and is the answer not in the manual?

Honeywell MultiRAE Series Specifications

General IconGeneral
BrandHoneywell
ModelMultiRAE Series
CategoryGas Detectors
LanguageEnglish

Related product manuals