Dynamic pump stall calibration is complete. You can now exit Diagnostic Mode.
Important! When you are done setting thresholds, exit Diagnostic Mode and test the instrument
before actual use.
Setting Pump Stall Threshold Values Static Method
Use the following values for reference when using the Static method to adjust the pump stall
threshold values:
Low Speed High Speed
Vacuum -2.5in Hg -10in Hg
Flow
rate
>200 cc/min >300 cc/min
Idle (I) 100 to 200 counts 150 to 250 counts
Block
(Block-Idle) > 100
counts
(Block-idle) > 100
counts
Stall
Setting
(Idle + Block)/2 (Idle + Block)/2
Honeywell MultiRAE Series™ 162 User Manual