Enter Programming in Basic Mode
A password is necessary for entry to Programming Mode in Basic Mode.
IMPORTANT! If you provide an incorrect password or if you skip the password, you can still
enter Programming mode, but you can then only perform calibration. You only have Select,
Back, and scrolling options.
1. To enter Programming Mode and have access to all programming features, press and hold
[MODE] and [N/-] until you see the Password screen.
2. Input the 4-digit password:
l
Increase the number from 0 through 9 by pressing [Y/+].
l
Step from digit to digit using [N/-].
l
Press [MODE] when you are done.
If you make a mistake, you can cycle through the digits by pressing [N/-] and then using
[Y/+] to change the number in each position.
Note: The default password is 0000.
Note: The password screen only appears when you enter the Programming Mode the first
time after turning the instrument on in Basic Mode. If you have input the correct password,
you do not have to input it again to enter Programming Mode until you turn the instrument
off and on again.
Once you enter Programming Mode, the Calibration menu is highlighted. Press [N/-] to step
through the programming screens.
Honeywell MultiRAE Series™ 69 User Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell MultiRAE Series and is the answer not in the manual?

Honeywell MultiRAE Series Specifications

General IconGeneral
BrandHoneywell
ModelMultiRAE Series
CategoryGas Detectors
LanguageEnglish

Related product manuals