Proper setting of the pump stall threshold is necessary so that if there is an obstruction to the
inlet, the pump will stop and the instrument will go into alarm. This prevents unwanted debris or
liquid from entering the pump and causing disruption or damage. The MultiRAE provides two
methods to set the pump stall threshold: Static and Dynamic.
Note: The Dynamic method uses an algorithm that takes external temperature into consideration
for greater accuracy.
If it is necessary to set the pump stall threshold, you must enter Diagnostic Mode. Also,
regardless of which method is used, when you are done setting thresholds, exit Diagnostic Mode
and test the instrument before actual use.
Entering Diagnostic Mode to Set the Pump Stall Threshold
1. Start with the instrument turned off. Hold down the [Y/+] and [MODE] keys until the
MultiRAE starts.
2. When you see the password screen, input your 4-digit password, and then press [MODE].
3. Once you have entered Diagnostic Mode, press [N/-] until the “Pump” screen is displayed.
4. Set the high and low threshold settings for pump stall using the instructions in this
section.
Honeywell MultiRAE Series™ 159 User Manual
CHAPTER
17
Pump Stall Threshold Adjustment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell MultiRAE Series and is the answer not in the manual?

Honeywell MultiRAE Series Specifications

General IconGeneral
BrandHoneywell
ModelMultiRAE Series
CategoryGas Detectors
LanguageEnglish

Related product manuals