EasyManuals Logo
Home>HP>Monitor>1332A

HP 1332A User Manual

HP 1332A
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Model 1332A
SEC'I'ION
V
PERFORMANCE CHECK AND ADJUSTMENTS
Performance Check
5-1. INTRODUCTION.
WARNING
u
Read the Safety Summary at the front
of this manual before installing or oper-
ating the instrument.
5-2. This section contains performance checks and
adjustment procedures. The performance checks
verify that the instrument meets its published spec-
ifications (table 1-1). Adjustment procedures are
provided to help maintain the instrument within spec-
ification limits.
5-3. RECOMMENDED 'TEST EQUIPMENT.
5-4.
Equipment required for the performance checks
and adjustment procedures is listed in table 5-1. Any
equipment that satisfies the critical characteristics
given in the table may be substituted for the recom-
mended
modells).
5-5. PERFORMANCE CHECK RECORD.
5-6. A Performance Check Record form is provided
at the end of this section for the purpose of re-
cording the results of the performance checks. This
form lists all of the performance checks and their
acceptable limits. The form may be removed from the
manual and retained as a permanent record of the in-
coming inspection or routine maintenance performed
on the instrument.
5-7.
PERFORMANCE CHECKS.
5-8. The performance checks are designed to verify
the published instrument specifications. Perform the
checks in the sequence given for best results.
5-9. During all tests, install BNC shorting caps (HP
Part No. 1250-0645) on all unused inputs. For options
with differential inputs, use the
+X-,
+Y-,
or
n
2-input connector unless indicated otherwise.
Standard instruments are shipped with-
out covers.
Operator protection must
be provided by the system in which the
instrument is used since dangerous
voltages are present in the instrument.
5-10. The position of a trace on the CRT always
refers to the center of the trace. The distance between
traces always refers to the orthogonal distance from
trace center; or, from the edge of one trace to the
corresponding edge of the other trace.
5-11.
DEFLECTION AMPLIFIER RISE TIME.
Rise
time, measured between 10% and 90% points on the
displayed waveform, shall be equal to or less than
the value shown in table 5-2 for both
X
and
Y
amplifiers.
Equipment Required:
Pulse Generator (HP Model 8012B)
Monitor Oscilloscope
Two
10:l Divider Probes
44-in. BNC Cable
5-12. Check deflection amplifier rise time as follows:
a. Connect equipment as shown in figure 5-1.
b. Set pulse generator output as follows:
Standard
.
.
. . .
.
.
.
.
.
.
. . .
.
. .
1
V
p-p, 100 kHz
Option 100.
. .
.
. . . . .
.
. .
.
.
. .
5
V
p-p, 100 kHz
Option 101.
.
.
. .
.
.
.
.
.
.
.
.
. .
10
V
p-p, 100 kHz
c.
Set
X
GAIN and
X
POSITION controls for full-
screen deflection with one dot on first graticule
line and other dot on last graticule line.
d. Set monitor oscilloscope to A-B MODE and
measure rise time between 10% and 90% points (dotted
lines on CRT).
e. Rise time shall be less than value shown in
table 5-2 for applicable instrument.
f. Disconnect pulse generator from X AMPLI-
FIER input connector and connect to
Y
AMPLIFIER
input connector.
g.
Disconnect probes from horizontal deflection
plates and connect to vertical deflection plates.
h. Repeat steps
b
through e.
i.
For option 106 repeat steps b through e for
-X
and
-Y
inputs.
j.
Disconnect test equipment.
Scans by ArtekMedia © 2008

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 1332A and is the answer not in the manual?

HP 1332A Specifications

General IconGeneral
BrandHP
Model1332A
CategoryMonitor
LanguageEnglish

Related product manuals