EasyManuals Logo
Home>HP>Monitor>1332A

HP 1332A User Manual

HP 1332A
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Performance Check
Model
1332A
b. Rotate
X
POSITION control fully ccw. Beam
should go off screen to left.
NOTE
If a capacitance bridge is used, special
care must be taken to null out capacitance
of connecting cable and to null out the
effect of the
1
megohm input resistance.
c. With no input to
Y
amplifier, rotate
Y
POSI-
TION control fully cw. Beam should move upward
and off screen.
d. Rotate
Y
POSITION control fully ccw. Beam
should move downward and off screen.
5-25.
DEFLECTION AMPLIFIER DYNAMIC RANGE.
The dynamic range shall extend to at least 1/2 screen
diameter beyond full screen.
5-21.
INPUT RESISTANCE.
Measure input resist-
ance of
X,
Y,
and
Z
amplifiers.
Equipment Required:
Equipment Required:
Pulse Generator (HP Model 8012B)
Monitor Oscilloscope
BNC Tee
Three 44-in. BNC Cables
5-22. Check input resistance as follows:
a. Measure each input resistance with ohmmeter
leads connected positive-negative and then
negative-
positive and take average of both readings.
5-26. Check deflection amplifier dynamic range as
follows:
b. Perform measurements listed in table 5-3.
a. Connect equipment as shown in figure 5-4.
NOTE
b.
Set pulse generator for positive 100-kHz out-
put with pulse width set for 50 percent duty cycle.
Model
1332A instrument having differ-
ential inputs, in addition to
low-im-
pedance input termination, contains a
termination resistor between each input
and ground.
c. Set monitor
oscilloscove trigger to internal and
--
sweep speed to display approximately one complete
cycle of waveform on
1332A CRT.
d. Set
133212
X
GAIN fully ccw.
Table
5-3.
Input Resistance
e. Adjust pulse generator output for 10-division
p-p deflection.
Option
110
Option
I
Option
21 0 21 1
Amplifier
Standard
X
axis
Y
axis
Z
axis
1
meg nominal
1 meg nominal
1 meg nominal
f. Using
1332A
X
POSITION control, position
baseline of displayed waveform to left side of screen
(Option 105: right side).
g. Increase pulse generator output amplitude by
50 percent. Shape of baseline portion of square wave
should not move horizontally or change shape.
5-23.
INPUT CAPACITANCE.
Capacitance of
X-,
Y-,
and Z-AMPLIFIER inputs must be 60 pF or less.
h. Change pulse generator polarity to negative
and repeat steps b through g except in step
f, position
base line to right side of screen (Option 105: left
side).
NOTE
For Options 110, 210, and 211 instruments
proceed to paragraph 5-25.
i. Interchange
X
and
Y
AMPLIFIER inputs.
Equipment Required:
j.
Set pulse generator output amplitude for
8-
division p-p deflection on 1332A.
LCR Meter
Test Leads
Adapter (HP Model
10110A)
k. Repeat steps b through h. In step f, for positive
pulse polarity, position baseline to lower edge of CRT
(Option 105: upper edge); for negative pulse pol-
arity, position baseline to upper edge of CRT (Op-
tion 105: lower edge).
5-24. Check input capacitance as follows:
a.
Using LCR meter, measure input capacitance
at
X,
Y,
and
Z
AMPLIFIER inputs. Capacitance shall
be 60
pF or less.
1. Disconnect test equipment.
Scans by ArtekMedia © 2008

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 1332A and is the answer not in the manual?

HP 1332A Specifications

General IconGeneral
BrandHP
Model1332A
CategoryMonitor
LanguageEnglish

Related product manuals