CALibration Subsystem
Copyright © Itech Electronic Co., Ltd. 108
Command Syntax
CALibrate:CURRent:METEr:POINt <point>
Parameters
P1 | P2 | P3 | P4
Examples
CAL:CURR:METEr:POIN P2
Related Commands
CAL:STAT CAL:SAV
CALibrate:CURRent:METEr[:LEVel]
This command is only used in calibration mode. It enters a calibration current
value that you obtain by reading an external meter. You must first select a
calibration level (with CAL:CURR:METE:POIN) for the value being entered.
These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE.
Command Syntax
CALibrate:CURRent:METEr[:LEVel] <NRf>
Parameters
<external reading>
Unit
A (amps)
Examples
CAL:CURR 3.2223
Related Commands
CAL:STAT CAL:SAV
CALibrate:VOLTage:POINt
This command can only be used in calibration mode. It is used to set the
calibration points of constanrd voltage mode. P1, P2 is used in low voltage
meter range, P3, P4 is used in high voltage meter range. It can use calibrate
voltage source and voltage meter.