SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 77
Related Commands
CURR:PROT:DEL
CURR:PROT:STAT
[SOURce:]CURRent:PROTection:DELay
This command specifies the time that the input current can exceed the
protection level before the input is turned off.
Command Syntax
[SOURce:]CURRent:PROTection:DELay <NRf+>
Parameters
0 to 60 seconds | MINimum | MAXimum | DEFault
Unit
seconds
*RST Value
3
Examples
CURR:PROT:DEL 5
Query Syntax
[SOURce:]CURRent:PROTection:DELay?
Returned Parameters
<NR1>
Related Commands
CURR:PROT
CURR:PROT:STAT
[SOURce:]CURRent:TRANsient:MODE
This command selects the operating mode of the transient generator as follows
in constant current mode.
CONTinuous: The transient generator puts out a continuous pulse stream
after receipt of a trigger.
PULSe: The transient generator puts out a single pulse upon receipt of a