IEEE-488 Command Reference
Copyright © Itech Electronic Co., Ltd. 26
Command Syntax
*ESE <NRf>
Parameters
0 to 255
Power-On Value
see *PSC
Examples
*ESE 129
Query Syntax
*ESE?
Returned Parameters
<NR1>
Related Commands
*ESR? *PSC *STB?
*ESR?
This query reads the Standard Event Status Event register. Reading the register
clears it. The bit configuration of this register is the same as the Standard Event
Status Enable register (see *ESE). See chapter “Programming the Status
Registers” for a detailed explanation of this register.
Query Syntax
*ESR?
Parameters
None
Returned Parameters
<NR1> (register value)
Related Commands
*CLS *ESE *ESE? *OPC