IEEE-488 Command Reference
Copyright © Itech Electronic Co., Ltd. 30
CURR:PROT:LEV MAX CURR:PROT:DEL 3
POW:CONF MAX POW:PROT:LEV MAX
POW:PORT:DEL 3 VOLT:ON:LATC ON
VOLT:ON MIN SENS:VOLT:RANG:AUTO ON
SENS:FUNC:CURR DC SENS:NPLC 7
SENS:AVER:COUN 8 SENS:AVER:STAT 1
TRIG:COUN 1 TRIG:DEL 0
TRAC:POIN 2 TRAC:FEED TWO
TRAC:FEED:MODE NEV
NOTE: *RST does not clear any of the status registers or the error queue, and
does not affect any interface error conditions.
*RST sets the trigger system to the Idle state.
*RST clears the presently active list.
Command Syntax
*RST
Parameters
None
Related Commands
*PSC
*SAV
*SAV
This command stores the present state of the electronic load to a specified
location in memory. Up to 100 states can be stored.
If a particular state is desired at power-on, it should be stored in location 0. It
then will be recalled at power-on if the power-on state is set to RCL0. Use *RCL
to retrieve instrument states.
NOTE: *SAV does not save the programmed trigger values
([SOURce:]CURRent:TRIGGer,[SOURce:]RESistance:TRIGGer,[SOURce:]VO
LTage:TRIGGer). Programming an *RCL or a *RST command causes the
triggered settings to revert to their [IMMediate] settings.
Command Syntax
*SAV <NRf>