SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 72
[SOURce:]CURRent
This command sets the current that the load will regulate when operating in
constant current mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate] <NRf+>
Parameters
0 through MAX | MINimum | MAXimum | DEFault
Unit
A (amperes)
*RST Value
MINimum
Examples
CURR 5 CURR:LEV 0.5
Query Syntax
[SOURce:]CURRent[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
CURR:RANG
[SOURce:]CURRent:RANGe
This command sets the current range of the electronic load module. There are
two current ranges.
High Range: model dependent, see Table 4-1
Low Range: model dependent, see Table 4-1
When you program a range value, the load automatically selects the range that
corresponds to the value that you program. If the value falls in a region where
ranges overlap, the load selects the range with the highest resolution.
NOTE: When this command is executed, the IMMediate, TRANsient,
TRIGgered, and SLEW current settings are adjusted as follows:
If the existing settings are within the new range: No adjustment is made.