Model DMM7510
7½ Digit Graphical Sampling Multimeter Calibration Manual Section 2:
Performance verification
DMM7510-905-01 Rev. A / April 2015 2-3
Recommended test equipment
The following table summarizes the recommended verification equipment. You can use alternate
equipment if that equipment has specifications that meet or exceed those listed in the table below.
Test equipment uncertainty adds to the uncertainty of each measurement. Generally, test equipment
uncertainty should be at least four times more accurate than corresponding Model DMM7510
specifications. The following table lists the uncertainties of the recommended test equipment.
Manufacturer Model Description Used for: Uncertainty
Fluke
5720A or 5730A
High Performance
Multifunction Calibrator
DCV, ACV, ACI, and
resistance
See Note.
Fluke
5725A
Amplifier
DCI and ACI
See Note.
Fluke
8508A
8.5 Digit Reference
Multimeter
DCV and resistance
See Note.
Keysight
Technologies
33250A
Function/Arbitrary
Waveform Generator
Frequency
See Note.
IET Labs, Inc.
1423-A
Precision Decade
Capacitor
Capacitance, 1 nF to
1 µF
See Note.
IET Labs, Inc.
HACS-Z-A-2E-1uF
Series HACS-Z High
Accuracy Decade
Capacitance Box
Capacitance, 1 µF to
100 µF
See Note.
Keithley
Instruments
8610 or 8620
4-Wire DMM Shorting
Plug
DCV, Digitize DCV, and
resistance
N/A
Caddock
Electronics
USF240-200k
200 kΩ, 0.01 %
tolerance, 2 PPM/ °C
resistor
10 µA and 100 µA DCI
See Note.
Vishay
CMF65
10 MΩ, 0.25 %
tolerance, 50 PPM/ °C
resistor
10 µA and 100 µA DCI
See Note.
HYMEG
Corporation
FA-65-1G and FA-
65-10G
1 GΩ and 10 GΩ, 2 %
tolerance, 25 PPM/ °C
resistor
910 M, parallel 1 G
and 10 G
See Note.
Refer to the manufacturer's specifications to calculate the uncertainty, which varies for each function
and range test point.
In this manual, the Model 8610 shorting plug is shown in the graphics. However, you can use either
the Model 8610 or Model 8620.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals