Section
4: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
4-10 DMM7510-905-01 Rev. A / April 2015
cal.adjust.count
This attribute returns the number of times the instrument has been adjusted.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Nonvolatile memory Not applicable
Usage
adjustments = cal.adjust.count
adjustments
The number of adjustments
Details
You can use this command if calibration is locked or unlocked.
Example
Count = cal.adjust.count
print(Count)
Assign the number of times the instrument has been adjusted
to a user variable named Count.
Output the value. Example output:
3
This shows that the instrument has been adjusted 3 times.
Also see
cal.adjust.date (on page 4-11)