2: Performance verification Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
2-32 DMM7510-905-01 Rev. A / April 2015
-- *********************************************
-- **** Display first reading from buffer ****
-- *********************************************
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Analog trigger")
num7 = buf.readings[1]
str7 = string.format('%3.4e', num7)
display.settext(display.TEXT2, str7) print(num7)
Frequency verification
To verify the Model DMM7510 frequency function:
1. Connect the Keysight Technologies 33250A function generator to the Model DMM7510 INPUT HI
and LO terminals, as shown in the following figure.
2. Set the function generator output impedance to High, amplitude of 5 V
RMS
, waveform to
square wave.
3. Enable the output.
4. On the Model DMM7510, press the FUNCTION key, select the Measure Functions tab, and
select Frequency.
5. Select the MENU key.
6. Select Measure Settings.
7. Set the Aperture to 250 ms.
8. Set the Threshold Range to 10 V.
9. Set the Threshold Level to 0 V.
10. Press the HOME key.
11. Source the voltage and frequency values as listed in Verify the frequency (on page 2-33
). For
each setting, be sure that the reading is within low and high limits.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals