7½ Digit Graphical Sampling Multimeter Calibration Manual  Section 4: 
 
DMM7510-905-01 Rev. A / April 2015  4-13 
 
cal.adjust.internaltemp 
This attribute returns the internal temperature of the instrument when calibration was last adjusted. 
Type  TSP-Link accessible  Affected by  Where saved  Default value 
Usage 
temperature = cal.adjust.internaltemp 
 
The internal temperature in Celsius (°C) 
Example 
 
print(cal.adjust.internaltemp) 
Returns the internal temperature of the instrument when 
calibration was last adjusted. 
Example output: 
Also see 
cal.adjust.tempdiff (on page 4-15) 
 
cal.adjust.rear.ac() 
This function starts the specified rear AC adjustment step. 
Type  TSP-Link accessible 
Affected by  Where saved  Default value 
   
Usage 
cal.adjust.rear.ac(step) 
cal.adjust.rear.ac(step, value) 
 
The rear AC adjustment step to start 
The value for this adjustment step (if applicable) 
Details 
The descriptions of these steps are: 
  AC Current Step 1, 0.5 A 1 kHz, AC current 10A range zero 
  AC Current Step 2, 2 A to 10 A 1 kHz, AC current 10A range full-scale 
For detail on how to use cal.adjust.rear.ac(), see Rear terminal adjustments (on page 3-6
). 
For code examples, refer to: 
•  Rear AC adjustment step 1: ACI 10 A Zero adjustment (on page 3-8) 
•  Rear AC adjustment step 2: ACI 10 A Full-scale adjustment (on page 3-9)