7½ Digit Graphical Sampling Multimeter Calibration Manual Section 4: TSP command re
ference
DMM7510-905-01 Rev. A / April 2015 4-19
cal.unlock()
This attribute unlocks calibration operations.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
cal.unlock(password)
password
A string containing the password to unlock calibration
Details
Calibration data is locked during normal operation. To perform calibration, you must unlock
calibration.
The default password is KI000CAL. You can use cal.password to change the default.
An error is generated if this command is issued when calibration is already unlocked.
Example
cal.unlock("KI000CAL")
-- Perform operations to generate the calibration data
cal.save()
cal.lock()
Unlock the calibration for the instrument using the default password.
Save the calibration data.
Lock the calibration data.
Also see
cal.lock() (on page 4-16)
cal.password (on page 4-17)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals