Section
2: Performance verification Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
2-40 DMM7510-905-01 Rev. A / April 2015
Range Nominal calibrator
value
(Ω)
Actual calibrator
value
(Ω)
Lower limit
(Ω)
Upper limit
(Ω)
100
0
-8.00E-03
8.00E-03
100.0000
9.9996340E01
9.997934E+01
1.000133E+02
1000
0
-8.00E-02
8.00E-02
1000.000
9.9998410E+02
9.997241E+02
1.000244E+03
1.0E+04
0
-8.00E-01
8.00E-01
1900.000
1.8999910E+03
1.898583E+03
1.901399E+03
Resistance verification
To check the resistance function, you will:
Use shielded, Teflon-insulated or equivalent cables in a 4-wire configuration
Characterize the calibrator 1 Ω, 10 Ω, and 909 MΩ nominal values with external reference DMM
Verify that the displayed readings fall within specified limits
For the 1 Ω, 10 Ω, and 1 GΩ ranges, verify accuracy from reference DMM readings
For the 100 Ω to 100 MΩ ranges, verify accuracy from actual calibrator source values
Characterize the calibrator and 909 MΩ resistor
If not already done, complete the procedure for characterizing the calibrator for dry circuit verification.
Refer to Characterize the calibrator for 1 Ω and 10 Ω (on page 2-38
).
To characterize the resistor:
1. Make connections between the resistor and the reference DMM as shown in the figure below.
Figure 4: 909 MΩ resistor characterization

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals