Model
DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual Section 2:
Performance verification
DMM7510-905-01 Rev. A / April 2015 2-63
Verify DC voltage zeros using the 4-wire short
To verify DC voltage:
1. Leave the short applied as described in Verify resistance zeros using a 4-wire short (on page 2-
62).
2. Select the DC Voltage function.
3. Set the range to 1000 V.
DC Voltage verification is from descending range order, starting with the 1000 V range and finishing
on the 1 V range.
4. Verify that the 1000 V range zero is within specification. See the table below.
5. Verify that the 100 V to 1 V range zero is within specification.
Verify DC Voltage zeroes
Range Lower limit Upper limit
5.00E-03
5.00E-04
1.20E-05
1 -2.00E-06 2.00E-06
Rear-panel verification
DC current 10 A range verification
To verify 10 A range:
1. Set the TERMINALS switch to REAR. Ensure that the orange R is displayed.
2. Press the FUNCTION key and select DC Current.
3. Press the HOME key.
4. Set the range to 10A.
5. Connect Model DMM7510, reference DMM, calibrator, and amplifier as shown in the following
figure.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals