EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Calibration Manual

Keithley DMM7510
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #97 background imageLoading...
Page #97 background image
Section
3: Calibration Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
3-20 DMM7510-905-01 Rev. A / April 2015
Save calibration and set the adjustment dates
Calibration is temporary unless you send the cal.save() command. Also, calibration data is not
saved if calibration is locked, if invalid data exists, or if all steps were not completed.
Use the following commands to save and lock calibration adjustments.
cal.save()
cal.lock()
reset()
Calibrations are complete after they have been saved and locked.
Example TSP calibration code
The following code examples include code for the calibration steps and code that displays prompts on
the Model DMM7510 front panel.
You may need to adjust the calibration commands in this code for your equipment setup.
Rear terminal adjustments
--Rear DC adjustment step 0: amplifier offset
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear() -- Clear the error event log and Home screen error triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #0, Mux Amp offset")
cal.adjust.rear.dc(0) -- Supports either 4-Wire short or open terminals.
code,message,severity = eventlog.next() -- Set variables to print the error
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #0, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals