4: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
4-12 DMM7510-905-01 Rev. A / April 2015
cal.adjust.dc()
This function starts the specified DC adjustment step.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
cal.adjust.dc(step)
cal.adjust.dc(step, value)
The DC adjustment step to start
The value for this adjustment step (if applicable)
Details
The descriptions of these steps are:
DC Step 0: Internal amplifier offset adjustment
DC Step 1: 4-wire zero for DCI 1 A to 3 A, DCV, Digitize DCV, 2W Ω, and 4W Ω
DC Step 2: Open Terminals:
• DCI, zero 10 µA to 100 mA
• DCV, 100 V and 1000 V full-scale
• Resistance, 1 Ω to 1 GΩ full-scale
• Digitize DCI and DCV full-scale
• Waveform, analog trigger
• Frequency, threshold level trigger
DC Step 3: 10 V, DCV 10 V range
DC Step 4: -10 V, DCV 10 V range
DC Step 5: 10 kΩ, Resistance, 4W 10 kΩ range
For detail on how to use cal.adjust.dc(), refer to DC voltage adjustment steps 0 to 5 (on page
3-
11).
This command generates an error if:
• Calibration is locked
• The step is out of sequence
• The step does not complete successfully
• The value that is passed is invalid for the step, out of range, or not needed
Also see
cal.adjust.ac() (on page 4-8)
cal.lock() (on page 4-16)
cal.unlock() (on page 4-19)