Section
4: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
4-12 DMM7510-905-01 Rev. A / April 2015
cal.adjust.dc()
This function starts the specified DC adjustment step.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
cal.adjust.dc(step)
cal.adjust.dc(step, value)
step
The DC adjustment step to start
value
The value for this adjustment step (if applicable)
Details
The descriptions of these steps are:
DC Step 0: Internal amplifier offset adjustment
DC Step 1: 4-wire zero for DCI 1 A to 3 A, DCV, Digitize DCV, 2W Ω, and 4W
DC Step 2: Open Terminals:
DCI, zero 10 µA to 100 mA
DCV, 100 V and 1000 V full-scale
Resistance, 1 Ω to 1 GΩ full-scale
Digitize DCI and DCV full-scale
Waveform, analog trigger
Frequency, threshold level trigger
DC Step 3: 10 V, DCV 10 V range
DC Step 4: -10 V, DCV 10 V range
DC Step 5: 10 kΩ, Resistance, 4W 10 kΩ range
For detail on how to use cal.adjust.dc(), refer to DC voltage adjustment steps 0 to 5 (on page
3-
11).
This command generates an error if:
Calibration is locked
The step is out of sequence
The step does not complete successfully
The value that is passed is invalid for the step, out of range, or not needed
Also see
cal.adjust.ac() (on page 4-8)
cal.lock() (on page 4-16)
cal.unlock() (on page 4-19)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals