4: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
4-8 DMM7510-905-01 Rev. A / April 2015
Example
acal.schedule(acal.ACTION_RUN, acal.INTERVAL_1DAY, 8)
Sets auto calibration to run every day at 8 am.
Also see
Auto calibration (on page 2-4)
acal.run() (on page 4-6)
cal.adjust.ac()
This function starts the specified AC adjustment step.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
cal.adjust.ac(step)
cal.adjust.ac(step, value)
The AC adjustment step to start
The value for this adjustment step (if applicable)
Details
The descriptions of these tests are:
• ACV adjustment (cal.adjust.ac())
ACV Step 1, 10 mV 1 kHz, ACV 100 mV range zero
ACV Step 2, 100 mV 1 kHz, ACV 100 mV range full-scale and 1 V range zero
ACV Step 3, 100 mV 50 kHz, ACV 100 mV frequency flatness adjust
ACV Step 4, 1 V 1 kHz, ACV 1 V range full-scale and 10 V range zero
ACV Step 5, 1 V 50 kHz, ACV 1 V range frequency flatness adjust
ACV Step 6, 10 V 1 kHz, ACV 10 V range full-scale and 100 V range zero
ACV Step 7, 10 V 50 kHz, ACV 10 V range frequency flatness adjust
ACV Step 8, 100 V 1 kHz, ACV 100 V range full-scale and 700 V range zero
ACV Step 9, 100 V 50 kHz, ACV 100 V range frequency flatness adjust
ACV Step 10, 700 V 1 kHz, ACV 700 V range full-scale