7½ Digit Graphical Sampling Multimeter Calibration Manual Section 2:
DMM7510-905-01 Rev. A / April 2015 2-61
Verify the capacitance
Description Verification point Lower limit (F) Upper limit (F)
10 % 1 nF range 1.0000E-10 9.7000E-11 1.0300E-10
70 % 1 nF range 7.0000E-10 6.9100E-10 7.0900E-10
10 % 1 µF range 1.0000E-07 9.8600E-08 1.0140E-07
100 %100 µF range 1.0000E-04 9.9500E-05 1.0050E-04