2.
The
Sweep
(horizontal) calibration coefficient is then the
period of the reference signal divided by the product of
the number of divisions used in step 1 for setup of the
reference
and the setting of the
SWEEP
TIME/DIV
control.
Using
the formula:
Sweep
coefficient
Period
of the reference signal (sec)
horizontal width (div) x
SWEEP
TIME/DIV
setting
3.
Remove the reference signal and input the unknown
sig-
nal,
adjusting the
SWEEP
TIME/DIV
control for
easy
ob-
servation.
Measure
the width of one
cycle
in divisions and use the
following relationship to calculate the actual period.
[EXAMPLE]
SWEEP
TIME/DIV
is 0.1 ms and apply 1.75 kHz reference
signal.
Adjust the
VARIABLE
so that the distance of one cy-
cle
is 5 divisions.
Substituting the given
value:
u
• * » «• -
+
1.75 (kHz)-i
Horizontal coefficient =
5
(div) x 0.1 (ms/div)
=
1.143
Then,
SWEEP
TIME/DIV
is 0.2 ms/div and horizontal am-
plitude is 7 divisions. (See Fig. 43)
Substituting the given
value:
Pulse
width = 7 (div) x 1.143 x 0.2 (ms/div) = 1.6 ms
Using
the formula:
Period
of unknown signal = Width of 1
cycle
(div) x sweep
coefficient x
SWEEP
TIME/DIV
setting
Fig.
43
13.
PULSE
JITTER
MEASUREMENT
1.
Apply the signal to the
INPUT
jack and set the vertical
MODE
to the channel to be
used.
Use
the
VOLTS/DIV
to adjust for an
easy
to observe
waveform
display.
Special
care
should be taken to ad-
just
the trigger group of controls for a stable display.
Set
the
SWEEP
VARIABLE
to CAL position.
2.
Set the
HORIZ
MODE switch to ALT, and depress the
HOLD
OFF control.
Adjust
the
DELAY
TIME
POSITION
for intensified display
of the waveform to be measured.
3.
Using the B
SWEEP
TIME/DIV
adjust the display for in-
tensificaion
of the entire
jitter
area
of the waveform.
4.
Set the
HORIZ
MODE to B.
Measure
the width of the
jitter
area.
The
jitter
time is this width in division multiplied by the
setting of the B
SWEEP
TIME/DIV
control.
Using
the formula:
Pulse
jitter
= Jitter width (div) x B
SWEEP
TIME/DIV
setting
[EXAMPLE]
The
example shows a
case
in which the
jitter
width was
measured
at 1.6 divisions wide with the B
SWEEP
TIME/DIV
set
at 0.2 fis. (See Fig. 44)
Substituting the given
value:
Pulse
jitter
= 1.6 (div) x 0.2
(/is)
= 0.32 /is
Pulse
jitter
Fig.
44
31
Adjusted
reference
signal-
Reference
signal
Fig.
42
Unknown
signal