D9U004GKC-0101
26
6.3 Environmental conditions(Graciella)
Conditions of Use
Ambient Temperature from 10 °C to +40 °C
Relative Humidity from 30 % to 75 %
Atmospheric Pressure 795 hPa - 1060 hPa
Storage and Transport Conditions
Ambient Temperature from - 10 °C to + 50 °C
Relative Humidity from 30 % to 75 %
Atmospheric Pressure 860 hPa to 1060 hPa
6.4 Electrical Specications (Graciella)
Voltage 100 V AC, 3,15 A
Voltage 110 V AC, 3,15 A
Voltage 120 V AC, 3,15 A
Voltage 127 V AC, 3,15 A
Voltage 230 V AC, 1,6 A
Frequency 50/60 Hz
Motor voltage 24 V AC
Protection against water penetration IPX4
Device protection class I
Classication of the included parts B
Maximum power input 230 V max. 1,6A; 100-127 V max.3,15A
Chair fuses 2xT1, 6 AL 250V, (version 230V), 2x T3, 15 AL
250V(version 100-127V)
6.5 Electromagnetic Compatibility
The chair is suitable for hospitals with the exception of nearby active RF surgical instruments and RF shielded rooms of
magnetic resonance systems, where the intensity of EM interference is high.
The chair has no necessary functionality dened.
WARNING!
The use of this device next to or in a block with other devices should be avoided, as this could cause incor-
rect operation. If such use is necessary, this instrument and other instruments should be monitored to verify
that they are operating normally.
List of used cables: network cable, maximum length 6 m
WARNING!
Use of accessories, transducers, and cables other than those specied or provided by the manufacturer of
this device could result in increased electromagnetic emissions or reduced electromagnetic immunity of this
device and cause improper operation.
WARNING!
A portable RF communication device (including terminal equipment such as antenna cables and external
antennas) should not be used closer than 30 cm (12 inches) from any part of the Graciella chair, including
cables specied by the manufacturer. Otherwise, the operation of this device may deteriorate.
WARNING!
Do not overload the chair over the permissible safe working load (SWL) and observe the motor loader (INT.)
in order to maintain the basic safety of the bed in terms of electromagnetic interference for the entire expect-
ed life of the chair.