6.5.1 Instructions and manufacturer's declaration - electromagnetic radiation
Radiation test Conformity
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic radiation
IEC 61000-3-2
Class A
Voltage uctuations / ickering emission
IEC 61000-3-3
Complying
6.5.2 Instructions and manufacturer's declaration - electromagnetic immunity
Endurance test Satisfactory level
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV for contact discharge
± 15 kV for air discharge
Near elds from RF wireless communication devices
IEC 61000-4-3
See Table 1
Electrical fast transient / burst
IEC 61000-4-4
±2 kV repetition frequency 100 kHz
Surge
IEC 61000-4-5
± 1 kV combined
± 2 kV between phase and ground
Conducted RF
IEC 61000-4-6
3 V (0,15 MHz – 80 MHz)
6 V ISM bands between 0,15 MHz and 80 MHz)
80 % AM at 1 kHz)
Power-frequency magnetic elds (50/60 Hz)
IEC 61000-4-8
30 A/m
Short-term voltage drop and voltage interruption (power supply)
IEC 61000-4-11
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° a 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles