61
13.9 Castors
The castors serve only for moving the chair within the surgery (eg. when cleaning).
WARNING! No one may sit on the chair when activating, using and deacti-
vating the castors.
► With the castors in the active position, the thresholds and other irregularities
must not be passed. Castors are not intended for transporting the chair.
► Always handle the chair with the mains cable disconnected.
► Ensure that no cables are run over when moving the chair.
► Disconnect the step before handling the chair.
Fig. Activating the castors
Fig. Deactivated castors
Fig. Active castors
Fig. Deactivating the castors
Procedure for activating / deactivating the castors
Activation of handling wheels:
1) Set the chair to the lowest position
2) Grasp the footrests and lift the chair slightly until you hear the sound (one click) of the
castors locking in the active position (if you hear 2 clicks, the castors are not in the active
position).
3) Lower the chair
4) When handling the chair, keep the chair in a horizontal position with footrests
Deactivation of the castors:
1) Grasp the footrests and lift the chair slightly until you hear the sound (one click) of un-
locking of the castors
2) Lower the chair