6 - 8 Operator’s Manual
6 Image Acquisition
1. Press <TDI> to enter the TDI mode.
– In B or B+Color mode: to enter TVI Mode, parameters of TVI mode will be displayed on
the touch screen.
– In Power mode: to enter TEI Mode, parameters of TEI mode will be displayed on the
touch screen.
– PW mode: Press <TDI> and then press <PW> or <Update> to enter TVD. The parameters
of TCD are displayed on the touch screen.
– M mode: Press <TDI> and then press <M> or <Update> to enter TVM. The parameters of
TVM are displayed on the touch screen.
2. Adjust the image parameters to obtain optimized images.
3. Press <TDI> to exit from TDI mode and enter general imaging modes.
Or, press <B> on the control panel to return to B mode.
6.4.2 TDI Mode Image Parameters
In each TDI mode, the parameters that can be adjusted are similar to those in the color flow modes
(Color, PW, and Power). See the relevant sections for details. The following introduces the specific
items in TDI mode.
Tissue State
This function is used for fast image optimization.
6.4.3 TDI Quantitative Analysis
CAUTION
TDI is provided for reference, not for confirming a diagnosis.
NOTE:
To perform the strain and strain curve, the ECG curve is in need in case of the deviation in curve.
TIP:
• TDI quantitative analysis is an option.
• The current image (in frozen state) and the saved image can be used in the quantitative
analysis.
• Only after the user chooses the image review, the quantitative analysis is available.If the user
chooses the static image (only one frame), the quantitative analysis is not available.
It is about analyzing the data of TVI imaging and measuring the velocity of the myocardium with
the cardiac cycle.
Here are three types of curves to perform the quantitative analysis:
• Velocity-time curve
• Strain-time curve
• Strain rate-time curve
– Strain: Deformation and displacement of the tissue within the specified time.

Table of Contents

Other manuals for Mindray ME7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME7 and is the answer not in the manual?

Mindray ME7 Specifications

General IconGeneral
BrandMindray
ModelME7
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals