EasyManuals Logo
Home>Mindray>Diagnostic Equipment>Nuewa R9 Elite

Mindray Nuewa R9 Elite User Manual

Mindray Nuewa R9 Elite
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Function and Performance Checking Method 5-11
Press <iStation> key to enter patient information
management (iStation page)
The saved patient information (images) can be
found, and the patient information can be:
Backed up/ Restored
Sent (To DICOM or USB disk etc. )
5.5 Performance Test
5.5.1 Test Process
Resolution
Detection depth
Geometric Position
Accuracy
Blackout area test
Record the exam
images
5.5.2 Test Content
NOTE:
The image used here is only for reference, stick to the image effect in the real
situation.
Requirements:
1. Display: set the contrast and brightness at the clinical application value (or the default status)
2. Operation environment: dark room, simulating the clinical application environment.
3. Scanning techniques: contact the probe with the acoustic window of the phantom, no spacing
nor pressing.
Tips:
For the testing phantoms, please refer to Appendix B.
KS107BD is low frequency phantom and used when Probe focus frequency is less than 4MHZ;
KS107BG is high frequency phantom and used when Probe focus frequency is more than
5MHZ;

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Nuewa R9 Elite and is the answer not in the manual?

Mindray Nuewa R9 Elite Specifications

General IconGeneral
BrandMindray
ModelNuewa R9 Elite
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals