Product Principle 4-5
4.2.3 Engine Board
Figure 5 Principle Diagram of Engine Board
The hardware circuit of engine board is shown in the figure above.
Function:
 Distribute the clock.
 Provides the control and management function of probe related.
a) Probe online, IDs, hot-plug.
b) Control of interior circuit of the probe, such as high-voltage switch, parallel turning, Cable
Driver, etc.
 Corresponding control signal of the probe board:
a) Control of the high-voltage (only applicable for the physical channel less than 128, Plato is
not equipped with the function).
b) Configuration of the CPLD on the probe.
c) SPI communicate with probe
 Responsible for the module resetting, online upgrade, the management of module information.
 Communication and control of the accessories related with image scan, the accessories
include:
a) 4D driving function module; it is an optional independent module, which belongs to the
engine board.
b) Physiological signals function module.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Nuewa R9 Elite and is the answer not in the manual?

Mindray Nuewa R9 Elite Specifications

General IconGeneral
BrandMindray
ModelNuewa R9 Elite
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals