3-8 System Installation
3.3.2 Device Transport
3.3.3 Fixing Methods
1. Press the button to lower the control panel to the minimum position when the
ultrasound device is powered on. Press the button
to push the moving mechanism
straight forward.
Protect the monitor with the red
foam, and use the rope to fasten it
around the device. Protect the
monitor from moving.
Lower the support arm
to the minimum
Fasten the moving
mechanism (from
moving). Tie the rope
around the back
handle. Do not press
the external cover of
the device hard.
Unlock the brake pedal
Four ends of the
ropes should be
fastened to four
directions of the
car. Keep the
device from
moving in the car