Calibration Procedure—
465B Service
Adjustment
Procedure
m.
CHECK—
Display
is
triggered along
the
negative
slope of
the waveform as
the B
TRIGGER LEVELcontrol
is
rotated.
19. Check Line
Triggers
a.
Set:
CH 1 VOLTS/DIV
and VAR As required
n. CHECK—
Display is not
triggered (not
visible) at
either
extreme of
rotation.
o.
Disconnect the test setup.
A
TRIGGER
MODE
AUTO
COUPLING AC
SOURCE
LINE
SLOPE
+
p.
Set:
HORIZ
DISPLAY
A
TRIGGER SOURCE
CH 1
VOLTS/DIV
CH 1 VAR
A
TRIGGER COUPLING
A
EXT/10
5
V
Fully
counterclockwise
AC
b.
Connect a 1
0X probe from the
CH 1 input to a
line-
frequency
source.
C.
CHECK—For a
stable
display, starting on the
positive-going
slope.
q.
Connect
the calibration
generator
standard-
amplitude output to
the CH 1
input via a 50 D
bnc cable
and bnc T.
Connect the
bnc female
connection to A
External Trigger
input via a 50
Q bnc cable.
Adjust the
generator output to 50
volts.
NOTE
The
range of the
A LEVEL
control with
the A
SOURCE switch
in EXT/10 is ±20
volts (40 V
peak-
to-peak) or
greater. The applied
signal
is 50 V
peak-
to-peak.
Therefore,
untriggered operation
at both
extremes of the A
TRIGGER LEVEL
control is not
required.
r.
CHECK—
Display is
triggered along the
negative
slope of
the waveform
as the A
TRIGGER
LEVELcontrol
is
rotated.
s. Set
the A
TRIGGER SLOPE
switch to +.
t. CHECK—
Display is
triggered along the
positive
slope of the
waveform asthe ATRIGGER
LEVELcontrol
is
rotated.
u.
Disconnect
the test
setup.
d. Set
the A
TRIGGER SOURCE
switch to
-.
e.
CHECK—For a
stable display,
starting on the
negative-going slope.
f. Disconnect
probe from the
line-frequency
source;
then from the 465B.
20. Check
NORM A Trigger
Mode
a.
Set:
A TIME/DIV
1 ms
CH 1 VOLTS/DIV
.5 V
CH 1 VAR
Calibrated detent
ATRIGGER
SLOPE
+
SOURCE
NORM
COUPLING
AC
MODE
AUTO
b.
Connect time-mark
generator to the CH 1
input via
a 50 O bnc
cable and 50 Q
bnc termination.
Adjust
generator for 0.1 s time
markers.