Calibration
Procedure—
465B
Service
Performance
Check
TRIGGERS
Equipment Required
1.
Leveled
Sine-wave
Generator 6. 50 Q Signal Pickoff Unit
(Type CT-3)
2.
Calibration Generator 7. 10X Bnc Attenuator
3. 42-inch Bnc
Cable
(2
required) 8. 2X Bnc Attenuator
4.
Gr-to-bnc-female Adapter 9. 50 fi Bnc
Termination
(2
required)
5.
Gr-to-bnc-male Adapter 10. Dual-input
Coupler
(2
required)
465B Control
Settings
POWER ON
CRT
INTENSITY As desired
FOCUS
Best focused
display
SCALE ILLUM As
desired
Vertical (CH
1 and CH
2)
VERT MODE CH
1
POSITION
Midrange
VOLTS/DIV 5
mV
VAR
VOLTS/DIV Calibrated
detent
AC-GND-DC
DC
INVERT
Normal (button out)
20 MHz
BW LIMIT Full
bandwidth
(button out)
Trigger (A and B)
COUPLING
AC
LEVEL
Midrange
SLOPE +
SOURCE
NORM
TRIG MODE AUTO
A
TRIGGER HOLDOFF NORM
Sweep (A and B)
HORIZ DISPLAY
A
TIME/DIV
5 /
js
VAR TIME/DIV
Calibrated detent
DELAY TIME POSITION
Fully counterclockwise
X10
MAG
Off (button out)
POSITION
(Horizontal) Midrange
1.
Check A and B Internal Triggering
a.
Connect
the output of the
leveled sine-wave
generator to the A and B External trigger inputs
via
a
50 Q
cable, Gr-to-bnc-male adapter, 50 D signal pickoff unit
(Type CT-3) THRU SIG OUT output,
Gr-to-bnc-female
adapter, 50 Q bnc cable, 2X bnc attenuator,
10X bnc
attenuator, 50 O bnc
termination
and
dual-input coupler.
b. Connect the SIG OUT 10% on the
CT-3 tot he CH 1
and CH 2 inputs via a 50 O bnc cable, 50 Ci bnc termina-
tion and dual-input coupler.
c.
Adjust the leveled sine-wave generator for
either
a
3-division,
50
kHz display and set both
VOLIS/DIV
switches
to 50
mV
(0.3
division of display), or a 5-division,
50
kHz display and set both
VOLTS/DIV switches to
50 mV
(0.5
division display) as necessary
for parts
d
and
f
of step 1.
d.
CHECK—Stable display can be obtained by
ad-
justing the A TRIGGER LEVEL
control
in
the following A
TRIGGER COUPLING
and A TRIGGER SOURCE switch
positions:
A
TRIGGER A TRIGGER
COUPLING SOURCE
DISPLAY
AC NORM, CH
1,
CH 2 0.3
Div
DC
NORM, CH
1,
CH 2 0.3
Div
LF REJ
NORM, CH
1,
CH 2 0.5 Div
HE
REJ NORM, CH
1,
CH 2 0.5 Div
e. Set:
HOHIZ
DISPLAY
B
DLY’D
A TRIGGER LEVEL Fully
clockwise