Calibration Procedure—
465B
Service
Performance
Check
c.
CHECK—Display
amplitude is 5 divisions ±5%
(5
divisions
±0.25 division).
d.
CHECK—
Display
will trigger
symmetrically
within
one
graticule division
when the A
TRIGGER
SLOPE is
switched
between
+ and
—
.
e. Set the A
TRIGGER SOURCEto EXT/IOand
change
the output of the calibration
generator to
5
V.
f.
CHECK—Display
amplitude is 5
divisions ±5%
(5
divisions
±0.25
division).
g.
Disconnect the
calibration
generator
standard-
amplitude
signal
from the A
External
Trigger input
connector.
18.
Check Trigger
View
Low-Frequency
Compensation
a. Set:
A
TRIGGER
SOURCE EXT
ATIME/DIV
0.1ms
b. Connect the
calibration generator
high-amplitude
output to the
A External
Trigger
input connector viaa50 O
bnc cable and 50 D bnc termination. Set the calibration
generator output for a vertical display of 5 divisions at
1 kHz.
c.
CHECK—Square
wave has less
than 10%
overshoot
or
rounding.
d. Set A
TRIGGER SOURCE to
EXT/10 and adjust
generator output for
a vertical
display of
5
divisions.
e.
CHECK—Square wave has less
than 10% overshoot
or rounding.
f. Disconnect the
calibration generator high-
amplitude output from
the
A
External Trigger input
connector.
19. Check Trigger View High-Frequency
Compensation
a. Set:
A TRIGGER SOURCE
EXT
A
TIME/DIV 0.2 ms
b.
Connect the calibration generator fast-rise + output
to
the
A
External Trigger input connector via a 50
0 bnc
cable,
and 50
O
bnc termination.
c.
Set the calibration generator
frequency
for 100
kHz
and adjust output for a vertical display
of
5
divisions.
d. CHECK
—Square-wave
front-corner aberration is
less
than ±10% (±0.5
division).
e.
Disconnect test setup.
20. Check Zero Trigger View
Delay
a. Set:
VERT MODE CH 1 and A TRIG
VIEW
ATIME/DIV 0.02 /us
XI0 MAG On
(button
in)
A TRIGGER
COUPLING AC
CH 1
VOLTS/D
IV
0.1
V
b. Connect output of sine-wave
generator via bnc T,
two 42-inch (equal length) 50 O bnc cables, and 50
O bnc
terminations; one to the CH 1 input connector,
and one to
the A External
Trigger
input connector.
Set sine-wave
generator for a
6-division display of A TRIG VIEW at
40 MHz, then adjust CH 1
VAR
and
VOLTS/DIV to match
the amplitude
of
the
CH 1 display totheamplitudeof theA
TRIG VIEW display.
c.
CHECK—Time difference
with displays superim-
posed is ^0.15
horizontal graticule divisions.
d. Set:
HORIZ
DISPLAY B DLY'D
ATIME/DIV 0.2
us
B
TIME/DIV 0.02
us
e CHECK—
Time difference with displays superim-
posed is ^0.15
horizontal graticule division.
f.
Disconnect test setup.