EasyManuals Logo
Home>Tektronix>Test Equipment>465B

Tektronix 465B User Manual

Tektronix 465B
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
Calibration
Procedure—465B Service
Adjustment Procedure
j.
INTERACTION—Between the two positions of the
DELAY TIME POSITION
dial when
adjusting C6015.
Compromise the
adjustments as necessary to achieve
correct timing.
12.
Check/Adjust B Sweep High Speed
Timing
(C6042)
a. Set:
DELAY TIME POSITION
Fully counterclockwise
HORIZ DISPLAY B DLY'D
B
TRIGGER SOURCE NORM
A
TIME/DIV
1 fjs
.
B
TIME/DIV 0.5 /us
b. Adjust the A and B
TRIGGER LEVEL controls for a
stable display.
c.
CHECK—Display for 1 time
marker/division
±2%
(1
minor division at the
eleventh time marker).
d.
ADJUST—B High Speed
Timing
(C6042).
with alow
capacitance screwdriver, for exactly 1 time
marker/divi-
sion.
13.
Check A and B Time/Division Accuracy
a. CHECK—Using the B
TIME/DIV switch and the
time-mark generator settings given in Table 4-15, check B
sweep timing is within 0.2 division, over
the
first 10
divisions of the display.
Table 4-15 (cont)
A and B
Time/Div Time-mark CRT
Display
Switch Generator (Markers/
Setting Output Division)
50 /us 50
/us 1
0.1 ms
0.1 ms 1
0.2 ms
0.2 ms 1
0.5 ms 0.5 ms
1
1 ms
1 ms 1
2 ms 2 ms
1
5 ms
5 ms 1
10 ms' 10 ms
1
20 ms* 20 ms
1
50
ms* 50 ms
1
A SWEEP ONLY
0.1 s* 0.1 s
1
0.2
s*
0.1 s 1
0.5
s*
0.5 s
1
‘Switch A TRIG MODE to
NORM below 5 ms/dlvision.
b. Set
the HORIZ DISPLAY switch
to
A.
c.
CHECK—Using the A TIME/DIV switch and the
time-mark generator settings given in
Table 4-15
check A
sweep timing within 0.2 division, over the first 10 divisions
of
the
display.
Table
4-15
A AND B
TIMING ACCURACY
A and B
Time/Div
Switch
Setting
Time-mark
Generator
Output
CRT Display
(Markers/
Division)
0.02
/us 20
ns 1
0.05 /us 50
ns 1
0.1 (is
0.1 (is 1
0.2
u
s
0.2 (is 1
0.5 /us 0.5
(is 1
1
/US 1
(IS
1
2
/us
2 us
1
5
(is 5 us
1
10
(IS
10 us
1
20
(is 20 /us
1
14.
Adjust
High-Speed
Magnified
Timing
(C4246,
C4247)
a. Set:
HORIZ
DISPLAY
X10 MAG
Horizontal
POSITION
A and B TIME/DIV
A
XI
0
(button
in)
Midrange (as required)
0.05 /us
b.
Select
5 ns time
markers
from the
time-mark
generator.
c. Set CH 1
VOLTS/DIV
for at least a 4-division
display.
d.
CHECK—Display
for
1
time
marker/division ±3%
(±1
1/2 minor divisions at the eleventh time marker).
@4-56

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 465B and is the answer not in the manual?

Tektronix 465B Specifications

General IconGeneral
BrandTektronix
Model465B
CategoryTest Equipment
LanguageEnglish

Related product manuals