Performance Tests
4- 104
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
H From the button bar touch Trig, select the A-- >B Seq tab, and touch
the A then B Trig After Time button.
H Select the BEventtab, and touch the Set 50% button.
H CHECK that a stable trigger is obtained for the test waveform for
both the positive and negative slopes of the waveform. Use the
TRIGGER LEVEL knob to stabilize the Main trigger. Touch B
Trig Level and use the keypad or the multipurpose knob/FINE
button to stabilize the Delayed trigger. Touch one of the Slope
buttons to switch between trigger slopes. See Figure 4--28 on
page 4--103.
H Leave the Delayed trigger system triggered on the positive slope of
the waveform before continuing to the next step. Also, return to the
main trigger: select the A-->B Seq tab and touch the A-->B Sequence
AOnlybutton. Then select the AEventtab.
H Press Close.
3. Confirm the AUX Trigger input (at 50 MHz):
a. Display the test signal:
H Remove the 10X attenuator and reconnect the cable to CH 1.
H Set the signal amplitude as follows:
CSA7404B 2.5 divisions
TDS7704B, CSA7404B, TDS7404B,
TDS7254B & TDS7154B 1.5 divisions
H Now fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is as follows (Readout may
fluctuate):
CSA7404B 250 mV
TDS7704B, CSA7404B, TDS7404B,
TDS7254B & TDS7154B 150 mV
b. Check the AUX trigger source for stable triggering at limits: Do the
following in the order listed.
H Use the definition for stable trigger from step 2b.
H Press the Trigger SOURCE button to toggle it to EXT.
H Press PUSH TO SET 50%.
H CHECK that a stable trigger is obtained for the test waveform on
both the positive and negative slopes. Press the Trigger SLOPE