EasyManuals Logo
Home>Tektronix>Test Equipment>TDS3000 Series

Tektronix TDS3000 Series Program Manual

Tektronix TDS3000 Series
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #341 background imageLoading...
Page #341 background image
Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 303
TRIGger:B:TYPe { EDGE }
TRIGger:B:TYPe?
EDGE is a normal trigger. A B trigger event occurs when a signal passes through a
specified voltage level in a specified direction and is controlled by the TRIG-
ger:B:EDGE commands.
TRIGGER:B:TYPE?
returns :TRIGGER:B:TYPE EDGE indicating that the B trigger type is a edge
trigger.
TRIGger:DELay Commands
The TRIGger:DELay commands listed in Table 2--42 have the same function and
syntax as the equivalent TRIGger:B commands, and are for compatibility with
other Tektronix instruments.
Table 2- 42: TRIGger:DELay commands
TRIGger:DELay TRIGger:DELay: BY
TRIGger:DELay: EDGE:COUPling TRIGger:DELay: EDGE:SLOpe
TRIGger:DELay: EDGE:SOUrce TRIGger:DELay:EVENTS:COUNt
TRIGger:DELay: LEVel TRIGger:DELay: SETLevel
TRIGger:DELay:STATE TRIGger:DELay:TIMe
TRIGger:DELay: TYPe
TRIGger:EXTERnal:PRObe
(TDS3014B/C, TDS3024B, TDS3034B/C, TDS3044B, TD S3054B /C, TDS3064B Only)
Sets or returns the gain factor value of the external probe connector.
Trigger
TRIGger:EXTERnal:PRObe <NR1>
TRIGger:EXTERnal:PRObe?
Syntax
Arguments
Examples
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals