Section 3 - Configuring the VM700A
VM700A Option 01 (NTSC) & Option 11 (PAL) User Manual
3-17
Reset Character: can be set to None or Ctrl-A through Ctrl-Z. (Note: the
Reset Character for an Apple LaserWriter is Ctrl-D.)
Carrier Detect: can be enabled or disabled.
The default Port 1 settings are suitable for an Epson LQ-570 printer.
To Change a Port Configuration Oarameter Value. Select the parameter you wish to
change by rotating the knob until the parameter is highlighted, then touching the
screen over the parameter value.
Rotate the control knob until the desired parameter value appears.
Press the Accept Input soft key, or
Touch the screen anywhere inside the window, then press the Update & Exit
soft key to save the changes.
When high-level diagnostics are run, any errors detected are written to the
Diagnostic Errors file. The contents of this file can be printed using the Print
File soft key, or the file can be deleted using the Delete soft key.
When diagnostic errors appear, refer to the appropriate servicing information to
locate the possible cause of the problem. (This information may be found in the
VM700A Service Manual.)
The maximum capacity for the Diagnostic Errors file is 100 lines in a circular
file (scrolling past the last entry returns to the top line of the file). If the number
of errors detected causes the file to exceed this length, the earlier results are
scrolled out of the file buffer. When this occurs, the Diagnostic Errors file
includes a message stating that some of the text has been scrolled out of the file.
The Diagnostics Selection file lists the diagnostics the VM700A can perform at
power up. If installed options require additional diagnostics, those are also listed.
As received from the factory, all diagnostics are performed by default. Any
diagnostic can be deselected by rotating the control knob until the diagnostic is
highlighted, then touching the right side of the highlighted line (above the words
Not Selected).
In the following example, the FilterBoard Diagnostic routine will not be
performed.
Diagnostic Errors
(File)
Diagnostics
Selection (File)

Table of Contents

Other manuals for Tektronix VM700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix VM700A and is the answer not in the manual?

Tektronix VM700A Specifications

General IconGeneral
BrandTektronix
ModelVM700A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals