10. Cleaning and Maintenance
116
I
PaX-i3D Smart
10.2.2
Maintenance Task Checklis
Maintenance Tasks
Maintenance
period
Before operation, ensure that the equipment is clean and ready for
use. Make sure that all parts which come into contact with the patient
have been cleaned.
Daily
After using the equipment, make sure that the main power switch has
been turned off.
Daily
Ensure that the equipment is firmly plugged into a dedicated power
source.
Daily
Ensure that the plug and power cord are not hot. Daily
Confirm that the orange (exposure) indicator lamp turns on when
the exposure switch is pressed. Ensure that the orange (exposure)
indicator light remains on for the entire duration of the exposure.
Daily
Ensure that the power cable is not kinked, broken, exposed and that it
is free of all other defects.
Daily
Confirm that activating the emergency stop switch ceases the unit’s
operation. Pressing the emergency stop switch should stop all
equipment movements and X-ray emission.
Weekly
Ensure that all visible labels are intact and legible. Weekly
Check for possible wear or damage to the exposure switch cable. Monthly
Confirm that the audio message is audible throughout the duration of
the exposure.
Monthly
Please refer to the “Constancy Test Requirement” for documents about the image
quality inspection methods.